Gilson 171/172 Diode Array Detectors

Gilson 171/172 Diode Array Detectors

 


With a remote flow cell, the power of the detector is located directly at the outlet of the column, which yields virtually no peak broadening. An interchangeable flow cell design allows for both 171 and 172 Detectors to accommodate analytical and semi-preparative to preparative applications.

 

 

Wavelength

172 DAD: 190 - 610 nm

171 DAD: 190 - 720 nm

Wavelength Accuracy

± 1 nm

Repeatability <0.1 nm

S/N Ratio

172 DAD: <2*10E-05 AU

171 DAD: <3*10E-05 AU

Detection Type (pixels)

172 DAD: 512-element photodiode array

171 DAD: 256-element photodiode array

Drift

Baseline Drift

<0.0005 AU/hour (254 nm, according to ASTM E685-93)

Wavelength Temperature Drift

172 DAD: <0.005 nm/°C

171 DAD: <0.006 nm/°C

Short Term Noise (signal/noise ratio)

172 DAD: <0.00002 AU (254 nm, according to ASTM E685-93)

171 DAD: <0.00003 AU (254 nm, according to ASTM E685-93)

Flow Cells

The following quartz flow cells are available:

Flow Cell

Pathlength

Volume

Pressure

Analytical

5 mm

12 µL

500 psi (34.5 bar, 3.45 MPa)

Preparative

0.05 mm

0.16 µL

500 psi (34.5 bar, 3.45 MPa)

Preparative

0.2 mm

0.7 µL

500 psi (34.5 bar, 3.45 MPa)

Light Source

Deuterium and tungsten lamps (2000 hours expected lamp life)

Remote Flow Cell

Optional

Resolution

172 DAD: 0.8 nm/pixel

171 DAD: 2.2 nm/pixel

Lamp Warm-up Time

15 – 30 minutes

Front Panel

POWER/STANDBY soft key and LED indicator lights for POWER, UV, VIS and ERROR

Contact Control

Two open collector outputs and two contact inputs

Software Control

Computer control via Ethernet communication and TRILUTION® LC Software

Data Acquisition

Between 5 and 10 points/sec with TRILUTION® LC Software

Environmental Conditions

For indoor use only

Altitude: 2000 m

Temperature: 5° – 40°C

Air pressure: 75 – 105 kPa

Pollution degree: 1 or 2, in accordance with IEC 66

Humidity: Maximum relative humidity 80% for temperatures up to 31°C, decreasing linearly to 50% relative humidity at 40°C